PCUS® pro Single

System Features

The PCUS®pro single-channel ultrasonic frontend is designed to meet requirements of high data acquisition rates and high inspection speeds. In combination with custom software modules, the user can inspect parts and components effectively at high quality levels.

The USB 2.0 interface (no power supply required) provides high-speed data transfers at a maximum of 40MB/sec. The small dimensions (120mm length, 100mm width and 40mm height) provide easy portability and near transducer setup.

System Characteristics

Transmitter

  • Single-channel transmitter

  • -50V to -250V pulse voltage, adjustable in 1V increments

  • Negative rectangular pulse

  • 50Ω output impedance

  • 0μs to 40μs pulse delay, adjustable in 5μs steps

  • 10ns to 500ns pulse width, adjustable in 2.5ns steps

  • Up to 2kHz pulse repetition frequency, depending on recording length, pulse width, sampling rate and transmitter voltage

 Receiver

  • Single-channel receiver

  • Pulse/Echo or Transmit/Receive mode

  • 500kHz to 30MHz (-3dB) frequency range

  • Up to four analog receiver band filters (user selectable frequencies)

  • TGC with 80dB dynamic range, adjustable in 0.1dB increments, 256 points and ±40dB/μs slope

  • Attenuation/Amplification >100dB, adjustable in 0.1dB increments

  • 100μVss input sensitivity

Signal Path

  • 0μs to 655μs transducer delay in 10ns increments

  • 65,535 samples maximum recording length

  • 14bit A/D conversion, maximum 100MS/sec

  • One start gate and four measurement gates

  • RF-data or compressed data recording

  • Rectification:  positive, negative or full-wave

 Interface and Connectors

  • Lemo 00 transducer connectors

  • USB 2.0 high-speed  USB-B connector
    max. 5V/500mA power consumption

  • Trigger IN:  TTL high/low active; pulse width >100ns, opto-coupled (MCX)

  • Trigger OUT:  LVTTK high active (MCX)

  • Optional 4-axis encoder interface, external connection

Software

  • Digitally signed drivers for Windows® XP (SP2 or higher), Windows® Vista and Windows® 7 (32bit and 64bit)

  • Managed Windows® API (based on .NET 4.0 framework)

 

The  PCUS®pro Single system meets all relevant requirements of DIN EN 12668, Part 1.

Leeb Hardness Tester

WRITE TO US

Quality Network Pvt Ltd.

Works

No. 10, II Cross Street, Kalaimagal Nagar,
Ekkattuthangal, Chennai - 600032 

CONTACT US

Registered Address

#38/GA, Shoreham, 5th Avenue, Besant Nagar, 
Chennai – 600 090

CONTACT DETAILS

Phone: +91 44 3551 0355

Emailqnet.india@gmail.cominfo@qnetindia.com

Visitors professional logo

Copyright © 2019 - 2020 Quality Network India Pvt Ltd.