PCUS® pro Single

System Features

The PCUS®pro single-channel ultrasonic frontend is designed to meet requirements of high data acquisition rates and high inspection speeds. In combination with custom software modules, the user can inspect parts and components effectively at high quality levels.

The USB 2.0 interface (no power supply required) provides high-speed data transfers at a maximum of 40MB/sec. The small dimensions (120mm length, 100mm width and 40mm height) provide easy portability and near transducer setup.

System Characteristics


  • Single-channel transmitter

  • -50V to -250V pulse voltage, adjustable in 1V increments

  • Negative rectangular pulse

  • 50Ω output impedance

  • 0μs to 40μs pulse delay, adjustable in 5μs steps

  • 10ns to 500ns pulse width, adjustable in 2.5ns steps

  • Up to 2kHz pulse repetition frequency, depending on recording length, pulse width, sampling rate and transmitter voltage


  • Single-channel receiver

  • Pulse/Echo or Transmit/Receive mode

  • 500kHz to 30MHz (-3dB) frequency range

  • Up to four analog receiver band filters (user selectable frequencies)

  • TGC with 80dB dynamic range, adjustable in 0.1dB increments, 256 points and ±40dB/μs slope

  • Attenuation/Amplification >100dB, adjustable in 0.1dB increments

  • 100μVss input sensitivity

Signal Path

  • 0μs to 655μs transducer delay in 10ns increments

  • 65,535 samples maximum recording length

  • 14bit A/D conversion, maximum 100MS/sec

  • One start gate and four measurement gates

  • RF-data or compressed data recording

  • Rectification:  positive, negative or full-wave

 Interface and Connectors

  • Lemo 00 transducer connectors

  • USB 2.0 high-speed  USB-B connector
    max. 5V/500mA power consumption

  • Trigger IN:  TTL high/low active; pulse width >100ns, opto-coupled (MCX)

  • Trigger OUT:  LVTTK high active (MCX)

  • Optional 4-axis encoder interface, external connection


  • Digitally signed drivers for Windows® XP (SP2 or higher), Windows® Vista and Windows® 7 (32bit and 64bit)

  • Managed Windows® API (based on .NET 4.0 framework)


The  PCUS®pro Single system meets all relevant requirements of DIN EN 12668, Part 1.

Leeb Hardness Tester



Quality Network Pvt Ltd.


Plot No. 09, CGE Colony, Thiruvanmiyur, 
Chennai – 600 041 

Registered Address

#38/GA, Shoreham, 5th Avenue, Besant Nagar, 
Chennai – 600 090

Phone: +91 44 2442 0043
Mob: +91 9444979858
Fax: +91 44 2442 0043



USA - Quality Network Inc.

11 Main Street
Sparta, NJ 07871-1979

Phone: +01 (973) 726-8399
Mobile: +01 (973) 699-4460

Email: md@qnetworld.com
Web: www.qnetworld.com

Germany - Q NET Engineering GmbH

Altenkesselerstrasse 17- B6
D-66115 Saarbrücken

Phone: +49 (681) 976-7153
Email: info@q-net.de
Web: www.q-net.de


Copyright © 2017 - 2018 Quality Network India Pvt Ltd.